9.
Contact quality monitor indicator for split neutral electrode
This indicator illuminates green if a split neutral electrode is connected and the contact resistance is
within an acceptable range. The indicator illuminates
red if the split neutral electrode is not connected
or not applied properly (e.g. bad contact quality or partly dislocated) or no neutral electrode is
connected (in both cases the activation of monopolar output is disabled).
10.
Contact quality monitor indicator for non-split neutral electrode
This indicator illuminates green if a non-split neutral electrode is connected.
11.
Neutral electrode socket
This socket connects the plug of a neutral electrode for monopolar application (applied part).
12.
UNIVERSAL socket
This socket connects the plug of an Olympus HF instrument with HF instrument recognition (applied
part).
7.022.211 / ISSUE 5 23 / 110 Chapter 1: PRODUCT SPECIFICATIONs