2-1 Visual inspection of the electrosurgical generator and accessories
1) Check that the labels according to chapter 5-1 (Front panel, Rear panel) are present and legible. The
product name should be clearly visible on the front panel. Verify that the type plate shows details about the
type of the device, reference number, line voltage range, supply frequency, output power classification,
output frequency, duty cycle, serial number, manufacturing date and manufacturer according to Fig. 5.2.1,
Fig. 5.2.2, Fig. 5.2.3 or Fig. 5.2.4
Fig. 5.2.1. Sample of type plate of the electrosurgical unit
Fig. 5.2.2. Sample of type plate of the electrosurgical unit (Standard PK version)
12345W13-678 2011-08
12345W13-678 2014-08
7.022.211 / ISSUE 5 76 / 110 Chapter 5: Inspection