(J012) with each other. In this way a ring communication is established.
2) The Error Message “E394 Communication error” should appear.
3) Record the test results in the Inspection Card (refer to chapter 5-3).
2-11-5 Restore of output power settings
Restore the settings by using the tables created in chapter 7-2-1. When not done while inspecting, restore all
power settings off the changed modes noted in table 7.2. Finally restore the last active procedure if applicable
and assign all modes to the sockets.
2-12 Inspection label (For applicable markets)
1) Cover or exchange an inspection label (as shown in Fig. 5.2.14) at the rear panel of the electrosurgical
unit’s housing and mark the due date of the next periodic safety check (month / year). The electrosurgical
unit must undergo a periodic safety check at annual intervals.
2) Record the due date in the safety check report (see chapter 5-3).
Fig. 5.2.13. Sample of an inspection label