ESG-400
3 Inspection Card
Product name:
REF no.:
Serial no.:
Software version:
Service / Maintenance
manual version:
Nr. Test type / mode Pwr.
level
Load
resistance
Requirement Measured
value
Test passed
Effect Yes No
Visual inspection of the electrosurgical unit and accessories
1 Labels N/A N/A Procedure passed N/A
2
Obstructions or damage of
unit and accessories
N/A N/A Procedure passed N/A
Contact quality monitor function
3 Any bipolar mode N/A N/A Procedure passed N/A
4
Any
monopolar
mode
Standby,
non-split
N/A 0 Ω, 5 Ω Procedure passed N/A
5 Standby, split N/A 140 Ω Procedure passed N/A
6 Standby, split N/A 170 Ω Procedure passed N/A
7
Activation,
split
N/A 170 Ω Procedure passed N/A
DC Resistance
8 DC resistance N/A N/A ≥ 2.0 MΩ
9
Earth resistance
(with power cord)
N/A N/A ≤ 0.3 Ω
Earth leakage current
10 Normal condition (NC) N/A N/A ≤ 0.5 mA
11 Single fault condition (SFC) N/A N/A ≤ 1.0 mA
7.022.211 / ISSUE 5 97 / 110 Chapter 5: Inspection