Message Cause and recommended action
THE SPECIFIED FILE NAME ALREADY
EXISTS.
PLEASE CHOOSE ANOTHER NAME.
The file name exists.
→Specify the file name other than the existing file
name.
THE SPECIFIED FOLDER NAME
ALREADY EXISTS.
PLEASE CHOOSE ANOTHER NAME.
The folder name exists.
→Specify the folder name other than the existing
folder name.
SD MEDIA IS LOCKED.
PLEASE UNLOCK.
The SDHC card is locked.
→Unlock the SDHC card you will use and insert
the SDHC card in this instrument.
LOW BATTERY.
PLEASE CONNECT TO THE AC
ADAPTER OR TURN OFF THE POWER.
The battery voltage is low.
→Stop all operations, such as image recording,
copying, deleting and formating the SDHC card,
etc., and immediately replace the battery. Or
connect the AC adapter.
HIGH TEMPERATURE (DISTAL END).
PLEASE IMMEDIATELY PULL OUT THE
INSERTION TUBE.
The self-check function was activated to prompt
termination of inspection because the distal end
became too hot.
→Immediately pull the insertion tube out of the
observation object.
TIP
This message will appear before the
atmospheric temperature of the distal
end reaches the maximum operating
ambient temperature.
HIGH TEMPERATURE (MAIN UNIT).
PLEASE TURN OFF THE POWER.
The self-check function was activated to prompt
termination of the inspection when the internal
temperature is too high.
→Stop inspection temporarily, allow this instrument
to cool down before turning the power back on
again.
OVER CURRENT (ARTICULATION
MOTOR).
PLEASE TURN OFF THE POWER.
The self-check function was activated to prompt
termination of operation because the insertion tube
is overloaded.
→Extend the insertion tube as much as possible
to reduce the loop amount, and turn power back
on without operating the [ANGLE/LOCK] joystick.
92