10-1 Operating environment
Item Specifications
Operating temperatures
Insertion tube In air : -25 to 100 °C
In water
:
10 to 30
Parts other than
insertion tube
In air : -10 to 40 °C (Battery-powered operation)
:
0 to 40 °C (AC adapter-powered operation)
:
0 to 40 °C (Battery charging in progress)
Operating atmospheric pressure
Insertion tube In air : Normal pressure (1,013 hPa)
In water :
IV94 series 1,013 to 1,368 hPa (up to water
depth 3.5m)
In water : IV96 series 1,013 to 2,026 hPa (up to water
depth 10m)
Parts other than
insertion tube
In air : Normal pressure (1,013 hPa)
Operating environment altitude
All parts Up to altitude 2,000 m
Operating environment humidity
All parts 15 to 90% (relative humidity)
Liquid resistance
Insertion tube No problem even if exposed to machine oil, light oil or 5%
saline solution.
Parts other than
insertion tube
Waterproofing
Insertion tube Waterproof construction. Can be used underwater with the
optical adapter attached.
The stereo measurement is not available underwater.
Parts other than
insertion tube
Water resistant proof construction. Use underwater is not
supported. This unit is not drip-proof when the battery
compartment door, other doors or caps are open.
Contamination level
All parts 2 (Including AC adapter)
NOTE
Meeting the operating environment specifications does not constitute any guarantee
against damage to or malfunction of this instrument.
Use AC adapter and battery charger indoors (normal temperature).
This instrument can be used outdoor if all the doors and caps are firmly closed when using
the battery.