Figure 2-8 Fastening hole location and specifications
To prevent equipment damage, apply the following instructions when attaching
accessories to the OmniScan SX or when mounting the OmniScan SX to any surface.
Use M4 × 0.7 mm screws with a nominal penetration length (into the instrument)
of 8 mm.
Make sure that air can circulate freely between the OmniScan SX’s heat sink and
the accessory or mounting surface.
When mounting the OmniScan SX to any surface:
Always use the two fastening holes.
Make sure that there is at least a 12 mm diameter surface contact area (see
Figure 2-9 on page 32) between the OmniScan SX and the supporting surface.