Technical Data
43
6.3.2 Current Outputs
The data designated "guaranteed" apply to a mains power supply of
230 VAC and to ohm resistive load (load flow = 1). The data designated
"typical" apply to inductive load.
Table 6-5:
Current outputs
For possible operational limits refer to sections 6.3.4, "Operational Limits in
Conjunction with Mains Supply" and 6.3.5, "Operational Limits with Current
and Voltage Amplifier in Parallel".
Current Outputs
1
Output currents
3-phase AC (L-N) 3 x 0 ... 32 A
1-phase AC (L-L)
2, 3
1 x 0 ... 32 A
1-phase AC (LL-LN)
2
1 x 0 ... 64 A
DC (LL-LN)
2
1 x 0..90A
Power
7
Typical Guaranteed
3-phase AC (L-N) 3 x 430 VA at 25 A 3 x 250 W at 20 A
1-phase AC (L-L)
2, 3
1 x 870 VA at 25 A 1 x 530 W at 20 A
1-phase AC (LL-LN)
2
1 x 500 VA at 40 A 1 x 350 W at 40 A
DC (LL-LN)
2
1x 700W at ±40A 1x500W at ±40A
Accuracy Typical Guaranteed
R
load
0.5 Ω
Error < 0.05 % rd.
4
+ 0.02% of rg.
Error < 0.15 % of rd.
+ 0.05% of rg.
R
load
> 0.5 Ω Error < 0.1 % of rg. Error < 0.3 % of rg.
Harmonic distortion
(THD+N)
5,6
0.05 % < 0.15 %
Phase error
5
0.05 ° < 0.2 °
DC offset current < 3 mA < 10 mA
Resolution 1
m
A, 2 mA (2 phases parallel), ...
Frequency range
9
0 1000 Hz
Trigger on overload Timer accuracy error < 1 ms
Short-circuit protection Unlimited
Open-circuit protection Open outputs (open-circuit) permitted
Connection 4 mm/0.16 " banana connectors, amplifier
connection socket
8
Insulation Reinforced insulation of power supply and all
SELV interfaces
Footnotes:
1.Data for three-phase
systems are valid for
symmetric conditions
(0 °, 120 °, 240 °) unless
specified otherwise.
2. For wiring of single-
phase modes see
chapter 7, "Increasing
the Output Power,
Operating Modes" on
page 65.
3.Single-phase mode
(in phase opposition).
4.rd. = reading;
rg. = range, whereat
n % of rg. means: n % of
upper range value.
5.Valid for sinusoidal
signals at 50/60 Hz and
R
load
0.5 Ω.
6.Values at 20 kHz
measurement
bandwidth, nominal
value, and nominal load.
7.Guaranteed data at
230 V mains for ohmic
loads (PF=1); typical
data for inductive loads.
Refer to the sections
about operational lmits
6.3.4. and 6.3.5.
8.For currents > 25 A, con-
nect test object only to
the 4 mm/0.16 " banana
connections and not to
the generator combina-
tion socket.
9.Current amplitude
derating at frequencies
above 380 Hz (see
Figure 6-4).

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Omicron CMC 353 and is the answer not in the manual?

Omicron CMC 353 Specifications

General IconGeneral
BrandOmicron