EasyManuals Logo

Omicron CMC 353 User Manual

Omicron CMC 353
100 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #45 background imageLoading...
Page #45 background image
Technical Data
45
Figure 6-3:
Typical compliance
voltage (50/60 Hz)
The high and low sensitivity curves in figure 6-3 correspond to the overload
detection sensitivity settings in the Test Universe software. The low
sensitivity curves show the maximum available peak compliance voltage,
which is mainly relevant for testing primary and electromechanical relays.
Figure 6-4:
Current derating at high
frequencies for sinusoidal
signals
30
40
50
60
70
80
0
10
20
0 5 10 15 20 25 30 35 40 45
5
0
Output current in A
rms
Compliance voltage in V
peak
1-phase low sensitivity 32 A (L-L)
1-phase high sensitivity 32 A (L-L)
1-phase low sensitivity 64 A (LL-LN)
1-phase high sensitivity 64 A (LL-LN)
1-phase high sensitivity 32 A (L-N)
1-phase low sensitivity 32 A (L-N)
0
4
8
12
16
20
24
28
32
0 100 200 300 400 500 600 700 800 900 1000
Frequency [Hz]
Max. Current [A
]
Frequency in Hz
Max. current in A