7.2 Single-Phase Operation of the CMC 353
7.2.1 1 x 32 A High Burden Mode (L-L)
1 x 0 ... 32 A (±45 A
DC
), max. 70 V
peak
, 1 x 870 VA at 25 A
The currents 1 and 2 of the current triple are phase-opposite. This doubles
the compliance voltage of a single output.
Observe the safety instructions given in Section 7.1 on page 65 when using
this operating mode.
Figure 7-1:
Single-phase operation,
1 x 32 A high burden mode
Refer to the output curves shown in the figures 6-1 through 6-5 in section
6.3.2, "Current Outputs" on page 43.
Load
1’
N’
NV
1
V
2
V’ = 2 x V