EasyManuals Logo

Omni 7000 Series User Manual

Omni 7000 Series
193 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #144 background imageLoading...
Page #144 background image
OMNI 4000/7000 Operations and Maintenance Guide Rev F
7-9
Maintenance
3.
Select an analog input calibration to verify
by following the instructions in Steps 3
through 8 in the Analog Inputs section in
Section 7.2.2 Calibration.
After you apply your chosen Calibration
Units, the Verify Input group will appear.
4.
a. Enter the Applied Value from the
calibration verification
instrumentation.
b. Click Verify Input.
5.
The Error field will now display the
percentage of error of the verification.
a. If the percentage of error is within
acceptable limits, you do not need to
re-calibrate the input channel.
b. If the percentage of error is not within
acceptable limits, you may need to
re-calibrate the input channel (go to
Section 7.2.2).
c. Click End Verification to stop the
process.
Analog Output Verification
To verify that the output calibration variables are correct or within the acceptable error tolerance,
follow these instructions:
1.
In the Actions ribbon, click Diagnostics.
2.
Click on Calibration Verification in the
Diagnostics tree.

Table of Contents

Other manuals for Omni 7000 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Omni 7000 Series and is the answer not in the manual?

Omni 7000 Series Specifications

General IconGeneral
BrandOmni
Model7000 Series
CategoryDesktop
LanguageEnglish

Related product manuals