74
SECTION 4
When SECS is Used
CIDRW System
User’s Manual
SECTION 4
Reading from/Writing to ID Tags
Subsystem command (GetStatus)
Subsystem command (PerformDiagnostics)
S18,F13 Subsystem Command Request (GetStatus) S,H→Ε,reply
L,3
1.<TARGETID> "00"-"31"
2.<SSCMD> "GetStatus"
3.L,0
S18,F14 Subsystem Command Acknowledge (GetStatus) S,HE
L,3
1.<TARGETID> "00"-"31"
2.<SSACK> "GetStatus"
3.L,s
1.<STATUS1>
s.<STATUSs>
When the specified target is invalid:
s = 0, SSACK = "CE" communications error
When SSCMD is invalid:
s = 0, SSACK = "CE" communications error
If the status of SSACK is other than "NO" (normal), the List of Status will comprise zero items.
S18,F13 Subsystem Command Request (PerformDiagnostics) S,HE,reply
L,3
1.<TARGETID> "00"-"31"
2.<SSCMD> "PerformDiagnostics"
3.L,0
S18,F14 Subsystem command Acknowledge (PerformDiagnostics) S,HE
L,3
1.<TARGETID> "00"-"31"
2.<SSACK>
3.L,s
1.<STATUS1>
s.<STATUSs>
When the specified target is invalid:
s = 0, SSACK = "CE" communications error
When SSCMD is invalid:
s = 0, SSACK = "CE" communications error
If the status of SSACK is other than "NO" (normal), the List of Status will comprise zero items.