95
CIDRW System
User’s Manual
SECTION 5
When SECS is Not Used
SECTION 5
Troubleshooting
Operation Check Flowchart
From Installation to Trial Operation
Errors are indicated by whether or not a response to the test command is received and by the status of
the Amplifier Unit indicators.
If the test command was received normally:
Indicators
RUN COMM NORM ERROR
Response code for the response
Type
Response
code
Function
Normal 00 The command was received normally.
Error occurrence
Test command transmission
RUN-LED is OFF
No
Yes
An error has occurred at the Amplifier Unit.
Amplifier Unit error Refer to page 96.
Check if the Amplifier Unit settings are correct.
If there is no response to the command: Refer to page 96.
Response received
Yes
Check the nature of the response.
If there is a response to the command
Refer to page 96.
No
(Lights once)