Service Manual
OPTIKON 2000
Cod. 111004SEN Rev.C
3-3
29) HANDPIECES AND OTHER NON DISPOSABLE ACCESSORIES ARE SUPPLIED NON
STERILE. THEY MUST BE FIRST CLEANED, THEN STERILIZED IN AUTOCLAVE AS
INDICATED ON THE RELATIVE INSTRUCTIONS. THEY COULD BE DAMAGED BY
THE USE OF OTHER STERILIZATION METHODS, LIKE HOT AIR OR "CHEMCLAVE"
STERILIZATION.
30) THE USE OF INFLAMMABLE ANAESTETICS, (N
2
0 ), ( 0
2
), SHOULD BE AVOIDED
UNLESS THEY ARE REMOVED BY A GAS ASPIRATION SYSTEM.
31) DO NOT ACTIVATE THE PHACO U/S HANDPIECE WITH THE PHACO TIP IN AIR.
THE ULTRASONIC POWER SHOULD BE APPLIED TO THE PHACO HANDPIECE WITH
THE PHACO TIP IMMERSED IN A TEST CHAMBER FILLED WITH INFUSION FLUID
OR IN A BEAKER OF STERILE FLUID AT AMBIENT TEMPERATURE. OTHERWISE
HANDPIECE DAMAGE MAY RESULT.
32) DO NOT CHECK VIBRATION BY PLACING HAND OR FINGER AGAINST HANDPIECE
TIP. PROLONGED EXPOSURE TO OR DIRECT CONTACT WITH THE VIBRATING TIP
MAY CAUSE DAMAGE TO HEALTHY TISSUE.
33) THE PHACO HANDPIECE MUST BE CONNECTED TO THE PULSAR
2
IRRIGATION-
ASPIRATION SYSTEM AND MUST NOT BE USED WITHOUT IT.
34) BEFORE USING A SURGICAL OR NON SURGICAL ACCESSORY IN CONNECTION
WITH PULSAR
2
, VERIFY THE COMPATIBILITY OF THE SAME WITH PULSAR
2
EQUIPMENT. COMPATIBILITY IS STATED ON THE ACCESSORY ACCOMPANYING
DOCUMENT.
35) DISPOSE WASTE COLLECTION FLUIDS ACCORDING TO THE LOCAL
LAWS/REGULATIONS ON DISPOSAL OF ORGANIC MATERIALS.
36) SOUND EMISSION CAPABILITY OF THE EQUIPMENT IS TESTED AT POWER-ON.
VERIFY THAT AN AUDIBLE BEEP IS EMITTED DURING SYSTEM BOOT.
37) A ROTATING LOGO, PLACED AT THE LOWER SIDE OF THE USER INTERFACE
SCREEN, SHOWS THAT THE SYSTEM IS RUNNING PROPERLY. IF SUCH SYMBOL
STOPS MOVING IT MEANS THAT THE SYSTEM IS IDLE AND EQUIPMENT
OPERATION MUST BE STOPPED.
38) THE “PC UPLINK” CONNECTION IS INTENDED FOR SERVICE PURPOSE ONLY. DO
NOT CONNECT ANY EXTERNAL DEVICE DURING SURGICAL USE OF THE
EQUIPMENT.