7. TROUBLESHOOTING
7.1 WARNING AND ERROR MESSAGES
The PULSAR
2
console delivers warning and error messages on the LCD to inform
the operator of situations which may require special attention.
A window pops up in the middle of the LCD giving full explanation of the
warning or error, if operation can continue, an OK button appears on the
window.
The operator must depress this button to recognize that the message has been
received. The pop up window will close and a small warning symbol will then
appear, as a reminder, on the warnings frame located on the right side of the
LCD. This reminder will automatically disappear when the cause of the warning
has been removed.
The table below summarizes the warning and error messages.
W
A
R
N
I
N
G
E
R
R
O
R
MESSAGE
1 X
Err 01: Serial comunication fault
2 X
Err 02: Low input pressure
3 X
Err 03: I/A cassette missing!
4 X
Err 04: I/A cassette full. Aspiration halted!
6 X
Err 06: Vacuum level error
8 X
Err 08: Diathermy power mismatch
9 X
Err 09: Air pressure*
10 X
Err 10: Low Ultrasounds power
11 X
Err 11: Vac offset
12 X
Err 12: Major leakage
13 X
Err 13: No irrigation
14 X
Err 14: Aspiration occluded
15 X
Err 15: Minor leakage
16 X
Weak handpiece
17 X
Priming interrupted by user
19 X
Err 19: Pressure level error
23 X
Err 23: Air enable failure*
24 X
Err 24: US enable/reading failure
25 X
Err 25: US enable failure
26 X
Err 26: Diath enable/reading failure
27 X
Err 27: Diathermy enable failure
29 X
Err 29: Watchdog failure
30 X
Err 30: Power supply/reference mismatch
* Used only when the expansion VIT MODULE unit for vitrectomy is connected.