2-3
PARAMETER SPECIFICATION
DIATH (DIATHERMY)
Type: ................................. bipolar generator - generator stops when RF
power is not needed
Operating frequency: .............. 2 MHz
Nominal power:..................... 7W (450 Ohm LOAD)
No load max. voltage: ............. 100 V
Default bipolar power: ............ user programmable
Available bipolar power: .......... 5 to 100% (percent).SURGEON MODE (linear
power): allows linear control of the
DIATHERMY power via system footswitch
depression
Handpiece type: .................... bipolar microforceps, slim stat pencil eraser,
intraocular diathermy pencils
Diathermy cable: ................... two poles, 26 gauges, 75 ohm, steam
autoclavable. Use only original OPTIKON 2000
S.P.A. diathermy cable
Control: .............................. system footswitch
PHACOEMULSIFIER
Handpiece type: .................... piezoelectric
Frequency: .......................... approx. 40KHz
Tip stroke: ........................... from 0 to 100µm
Power: ............................... panel or linear control of the U/S power via
system footswitch depression.
U/S Mode: ........................... Linear or Panel; Continuous, Short Pulse,
Single Burst, Multi-Burst, Continuous Burst
U/S Timer: .......................... from 0.00 minutes to 9.59 minutes – Equivalent
Phaco Time display