Installation Data Oxford Instruments Plasma Technology PlasmalabSystem100
Installation Data (PECVD - TEOS)
Printed: 10-Feb-11, 9:07 Page 23 of 24 Issue 10: February 11
8.2.2 System Link Configuration Table
NAME
FUNCTION
NOTES
LK1
ANALOGUE 0V TO CHASSIS
LK2
DIGITAL 0V TO ANALOGUE 0V
LK3
NON CONTROLLER CRYO ENABLE
LK4
HEATER SNAP SWITCH BYPASS
LK5
FIT IF NO OEM CONTROLLER
LK6 A/B
LK6A = NON PM140 ENDPOINT
LK6B = PM140 ENDPOINT
SEE LK7 A/B
LK7 A/B
LK7A = NON PM140 ENDPOINT
LK7B = PM140 ENDPOINT
SEE LK 6 A/B
LK8
+24V DC TO RL2 COM1
LK9
+24V DC TO RL2 COM2
LK10
+24V DC TO RL6 COM1
LK11
+24V DC TO RL6 COM2
LK12
+24V DC TO RL7 COM1
LK13
+24V DC TO RL7 COM2
LK14 A/B
LK14A = NON DRY PUMP FITTED
LK14B = DRY PUMP FITTED
LK15
PUMP CURRENT BYPASS
LK16
+24V DC TO RL8 COM2
LK17
NON CONTROLLER HEATER
ALARM
LK18
NON CONTROLLER HEATER
ENABLE
LK19
SPARE INT/LOCK 1 BYPASS
NOT FITTED 100/133
LK20
SPARE INT/LOCK 2 BYPASS
LK21 A/B
LK21A = PURGE SWITCH NOT
FITTED
LK21B = PURGE SWITCH FITTED
Table 3: System link configuration