EasyManua.ls Logo

Oxford Instruments PlasmalabSystem100 - Heater and Chiller Specifications; Chiller Model Specifications and Performance

Oxford Instruments PlasmalabSystem100
24 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
PlasmalabSystem100 Oxford Instruments Plasma Technology Installation Data
Installation Data (PECVD - TEOS)
Issue 10: February 11 Page 18 of 24 Printed: 10-Feb-11, 9:07
7. Heater/chiller information
Chiller
Width
mm
Length
mm
Height
mm
Weight
kg
Usable
Temperature
Range °C
Water
Connections
Electrical
Requirements
Julabo FP51
460
550
890
90
-30 to 80
3/8”
Swagelok
400V, 50Hz,
3 Phase,
14 Amps
Julabo FP51
460
550
890
90
-30 to 80
3/8”
Swagelok
230V, 60Hz,
3 Phase,
14 Amps
Julabo FP51
460
550
890
90
-30 to 80
3/8”
Swagelok
220V, 60Hz,
1 Phase,
14 amps
Julabo FC 1600T
460
610
490
67
0 to 80
3/8”
Swagelok
230V, 50Hz,
1 Phase,
13 Amps
Julabo FC 1600T
460
610
490
67
0 to 80
3/8”
Swagelok
230V, 60Hz,
1 Phase,
13 Amps
Julabo FL 601
320
500
600
48
0 to 30
3/8”
Swagelok
230V, 50Hz,
1 Phase,
5 Amps
Julabo FL 601
320
500
600
48
0 to 30
3/8”
Swagelok
230V, 60Hz,
1 Phase,
5 Amps
Neslab
Thermoflex TF
1400
361
625
694
60
5 to 80
3/8”
Swagelok
208V, 60Hz,
1 Phase,
13 Amps
IMPORTANT: Heater/chillers should be filled/topped up with one of the fluids listed I
the following table, as appropriate. Refer the OIPT ‘Services Specifications
for Plasmalab and Ionfab systems’ document, sub-section 2.1 for the
warranty impact of not using these products.