Sensitivity Detection Procedures (Common to all Models)
1. Set the adjustment mode. (Refer to [LISTA Adjustment Modes] of the Table of Test Point.)
2. Playback the LISTA Alignment Tape.
3. Select the item <1> Sensitivity Measurement on the LISTA main menu, and then appeared 1.2% Speed. . .
on the screen, press the Enter key, and then start measurement of the sensitivity value.
4. Confirm the sensitivity value is within specification, when the message <<Sensitivity Measurement Finish>>
and Sensitivity = Sensitivity Value are displayed on screen.
5. If it is out of specification, repeat the step item 3 and 4.
6. If still out of specification, execute the Sensitivity Adjustment again.
Confirm this value