2 FUNCTIONS AND OPERATIONS
WARNING
g If the patient does not stay in bed without moving for 5 seconds or more
afterreturningtothebed,detectionmaynotwork,noticationmaynotbe
made,andthepatientmayfalloorfallover,resultingininjury.
g If the patient does not stay in bed without moving for 5 seconds or more
afterbedoperation,detectionmaynotwork,noticationmaynotbemade,
andthepatientmayfalloorfallover,resultingininjury.
g Set the weight according to the weight of the patient. If the setting is not
correct, detection may not work.
g This bed is designed for use on patients weighing equal to or more than
40 kg. Do not use BED EXIT ALARM on patients less than 40 kg.
Note!
<Falsenotication>
nIfthenoticationfunctionsarenotsetcorrectly,anaccidentalnoticationmaybemade.
nNoticationmaybemadeinthefollowingcases.Insuchacase,turnoorpausenotica-
tion.
• If a person other than the patient gets into or out of the bed
• If the patient rolls over to the foot end of the bed
• If an object of 3 kg or heavier is removed from the bed
• If the board is detached from the bed
• If the weight is applied to the footboard
n The larger the angle of the head section is, the higher is the possibility that the SIT-UP no-
ticationfunctionwilldetectaslightmovementsuchasrollingoverandwillmakenotica-
tion, even when the patient does not sit up.