Symptoms and status Check Corrective action
Reference
page
Noticationisnotmade
even though, for example,
the patient is out of bed
(type with BED EXIT
ALARM only).
Is the power plug disconnected?
Insert the power plug into the outlet. If the power plug is
not inserted into the outlet, you cannot use BED EXIT
ALARM.
41
Is the relay unit connected with the cable proper-
ly?
Connect the cable properly while referring to the instruc-
tion manual for the relay unit.
Is the bed communication cable disconnected
from the bed or the nurse call wall plate?
Firmly connect the bed communication cable. 48
Are the nurse call system and bed communica-
tion cable suitable for the bed?
For details on the suitable nurse call system and suitable
bed communication cable, contact Paramount Bed (see
p. 118) or your distributor.
Isthenoticationlampontherelayunitlit? Turnonnoticationontherelayunit. *
Is the power lamp of the BED Navi (accessory)
ashingorange?
Check the error code and then take corrective action. 99
Are there any obstacles between the bed and
oor?
Remove any objects.
Is the weight of the patient 40 kg or more?
BED EXIT ALARM is intended to be used for a user of at
least 40 kg, so a user less than 40 kg may not be detect-
ed.
*
Is the weight set according to the weight of the
patient?
Set the weight according to the weight of the patient.
Arethenoticationsettingsconguredproperly? Congurethenoticationsettingsproperly. *
Is the “MONITORING” function selected?
If “MONITORING”is selected, notication is not made
when the patient gets out of the bed, sits down on the
edge of the bed, or sits up.
56
Did the patient take an action such as getting out
of the bed during bed operation?
Noticationisnotmadeduringbedoperationevenwhen
the patient gets out of the bed or sits up. If you consider
that the patient may get into or out of the bed during bed
operation, set Overall lock-out.
65
Did the patient get out of the bed or sit up slowly?
If it takes 4 minutes or more for the patient to get out of
thebedorsitup,noticationmaynotbemade.
54
Did the patient get out of the bed or sit up after a
heavy object was placed on the bed?
If the patient gets out of the bed or sits up on the bed
after a heavy object (which weighs more than half the
patient’sweight)wasplacedonthebed,noticationmay
not be made.
Notication is not made
even though the patient
sat up (type with BED
EXIT ALARM only).
Isthepatient’ssit-upmotionsignicantenough?
Ifthepatient’supperbodymotionisinsignicant,noti-
cationmaynotbemade.Thisisnotafailure.Ifnotica-
tion is not made even though the patient raises the upper
bodytosomeextent,turno󰀨noticationonceandthen
turn it on again.
*
Did the patient sit up while his or her head is po-
sitioned at the foot end of the bed?
If the patient sat up while his or her head is positioned at
the foot end of the bed, detection may not work. After
getting the patient to lie down with his or her head posi-
tioned at the head end of the bed, turn o󰀨 notication
once and then turn it on again.
54 and
*
Did the patient sit up by propping themselves up
with a hand or elbow at the head end of the bed?
If the patient sits up by propping themselves up with a
handor elbowat the head end of the bed,notication
may not be made.
54
*) For the settings of BED EXIT ALARM, see “Connecting the bed communication cable” (p. 48) or the
instruction manual for the relay unit.