Symptoms and status Check Corrective action
Reference
page
Notication is made even
though the patient did not
perform actions such as
getting out of the bed (type
with BED EXIT ALARM
only).
Arethenoticationsettingsconguredproperly? Congurethenoticationsettingsproperly. *
Is an object of 3 kg or more removed from the
bed?
If an object of 3 kg or more is removed from the bed,
notication may be made. In such a case, turn o or
pausenotication.
55 and
*
Did a person other than the patient get into or out
of the bed?
If a physician, nurse, or visitor gets on or out of the bed,
notication may be made. In such a case, turn o or
pausenotication.
Is a board detached?
Ifaboardisdetached,noticationmaybemade.Insuch
acase,turnoorpausenotication.
Is weight applied to the footboard?
Ifweightisappliedtothefootboard,noticationmaybe
made.Insuchacase,turnoorpausenotication.
Is the angle of the head section large?
The larger the angle of the head section is, the higher is
the possibility that the SIT-UP detection function makes
noticationwithaslightmovementeventhoughthepa-
tient does not sit up.
55
Noticationismadeearlier
thanthespeciedmonitor-
ing time.
Was the monitoring time setting changed while
the patient is out of bed?
Turn o notication and then set the monitoring time
again while the patient is lying on the bed.
*
Charging of the USB de-
vice does not start or
charging stops premature-
ly.
Is the power plug inserted into the outlet? Insert the power plug into the outlet. 41
Is charging with the AC adapter for USB devices
possible?
Thecharge currentmaybeinsucientorthebedmay
not be suitable for the USB device. Disconnect the USB
device from the USB port and charge it with an AC
adapter for USB devices.
―
Although the USB device
is being charged, the re-
maining battery power
does not increase or
charging takes a long
time.
Does the USB device operate?
Thecharge currentmaybeinsucientorthebedmay
not be suitable for the USB device. Disconnect the USB
device from the USB port and charge it with an AC
adapter for USB devices.
―
*) For the settings of BED EXIT ALARM, see “Connecting the bed communication cable” (p. 48) or the
instruction manual for the relay unit.
If you request repair or if you have any questions, please provide us with the number displayed on the
product ID label.