F Disinfecting Your Nebuliser Handset
16 eFlow
®
rapid nebuliser system – Questions about your device? Contact your service partner
Method 2: Steam Disinfection
Step F5 Air-dry parts on a clean, lint-free towel.
Fig. F5
Step F6 After the parts are completely dry, store them in a
clean bag or cover them with a clean, lint-free towel.
Reassemble just before treatment.
Steam disinfection can be performed with a commercially
available baby bottle disinfector with an operating time of at least
6 minutes.
Step F7 Follow the directions for the baby bottle disinfector
and use the full disinfection cycle with the correct
volume of water described in the directions.
Fig. F7
Step F8 Remove the parts after the disinfector has cooled
and shake the parts to remove excess water.
Step F9 Air-dry parts on a clean, lint-free towel.
Step F10 After the parts are completely dry, store them in a
clean bag or cover them with a clean, lint-free towel.
Reassemble just before treatment.