K Troubleshooting
30 eFlow
®
rapid nebuliser system – Questions about your device? Contact your service partner
Controller Signal When Failures Occur and Action is Required
5. Flashing
alternating signal
tone
Normal: The device was
turned off.
If the device was switched off
manually and medication
remains, restart the controller.
6. Steady Normal: The easycare mode
was selected.
No action required. The
easycare is connected and
working properly.
# Controller Screen
Display
Display Symbol/
Controller Sound
Condition Action Required
# Controller Screen
Display
Display Symbol/
Controller Sound
Condition Action Required
7. Flashing
alternating signal
tone
• Bad or missing connection. • Ensure, that all parts are
connected tightly.
• Replace aerosol head and/or
the connection cord, if
defective.
• Restart treatment by pressing
ON/OFF button.
• If the condition persists,
contact your service partner.
8. Flashing
alternating signal
tone
• No drug detected. • Add drug, if necessary.
• Hold the nebuliser handset
level and restart treatment by
pressing the ON/OFF button.