K Troubleshooting
30 eFlow
®
rapid nebuliser system – Questions about your device? Contact your service partner
Controller Signal When Failures Occur and Action is Required
5. Flashing
alternating signal
tone
Normal: The device was
turned off.
If the device was switched off
manually and medication
remains, restart the controller.
6. Steady Normal: The easycare mode
was selected.
No action required. The
easycare is connected and
working properly.
# Controller Screen
Display
Display Symbol/
Controller Sound
Condition Action Required
# Controller Screen
Display
Display Symbol/
Controller Sound
Condition Action Required
7. Flashing
alternating signal
tone
Bad or missing connection. Ensure, that all parts are
connected tightly.
Replace aerosol head and/or
the connection cord, if
defective.
Restart treatment by pressing
ON/OFF button.
If the condition persists,
contact your service partner.
8. Flashing
alternating signal
tone
No drug detected. Add drug, if necessary.
Hold the nebuliser handset
level and restart treatment by
pressing the ON/OFF button.