EN
eFlow
®
rapid nebuliser system – Questions about your device? Contact your service partner 31
en
K Troubleshooting
9. Flashing
alternating signal
tone
The device has reached
20-minute maximum time
and will shut off.
Check to see if medication
remains in the medication
reservoir (see Notes, Step I9).
If medication remains, restart
the controller.
After treatment is complete,
clean the aerosol head.
10. Flashing
alternating signal
tone
Battery is empty, no
nebulisation is possible.
Connect power supply to take /
resume treatment and keep
power supply plugged
afterwards to charge battery.
11. Flashing
alternating signal
tone
Battery pack is defective. Switch to power supply to
continue treatment.
Contact your service partner.
12. Flashing
alternating signal
tone
Battery pack is overheated. Wait until the device cools
down or switch to power
supply to continue treatment.
# Controller Screen
Display
Display Symbol/
Controller Sound
Condition Action Required