rapid nebuliser system – Questions about your device? Contact your service partner 31
en
K Troubleshooting
9. Flashing
alternating signal
tone
• The device has reached
20-minute maximum time
and will shut off.
• Check to see if medication
remains in the medication
reservoir (see Notes, Step I9).
• If medication remains, restart
the controller.
• After treatment is complete,
clean the aerosol head.
10. Flashing
alternating signal
tone
• Battery is empty, no
nebulisation is possible.
• Connect power supply to take /
resume treatment and keep
power supply plugged
afterwards to charge battery.
11. Flashing
alternating signal
tone
• Battery pack is defective. • Switch to power supply to
continue treatment.
• Contact your service partner.
12. Flashing
alternating signal
tone
• Battery pack is overheated. • Wait until the device cools
down or switch to power
supply to continue treatment.
# Controller Screen
Display
Display Symbol/
Controller Sound
Condition Action Required