K3.1.142e Manual NitroFlow Lab - 16 -
4 Technical specifications
4.1 General
Delivery pressure
Maximum delivery pressure 8 bar(g)/116 psig
Ambient conditions
Temperature 10 to 35 °C / 50 to 95 °F
Air quality Normal clean ambient air, relative
humidity < 90%
Max. ambient relative humidity <80 % to 31°C, 50% at 40°C
Noise level < 65 dB(A) @ 1 meter/3 ft
Heat dissipation 1000 W (@50 Hz), 1200 W (@60
Hz) 3850 (50Hz)/4625 (60 Hz)
BTU/hr
Dimensions and connections
Dimensions (H x W x D) [mm] 700 x 900 x 310
Dimensions (H x W x D) [inch] 27.6 x 35.4 x 12.2
Net weight 92.5 kg / 204 lbs
Connections outlet: G ¼ “ / ¼” NPT
Electrical data
Voltage/frequency
1
120Vac/60Hz, 230Vac/50Hz
Power consumption 1400 W
Plug/receptacle 230V/50Hz can be mounted locally
as to local requirements
115V/60Hz: 20A/125 VAC; NEMA 5-
20 straight blade
1
Mains supply voltage fluctuations not to exceed +/- 10% of nominal voltage.