K3.1.142e Manual NitroFlow Lab - 20 -
5 Installation
Follow the paragraphs in this chapter to install the NitroFlow
®
Lab.
5.1 Transport
Warning
Transport the NitroFlow
®
Lab upright.
Put the NitroFlow
®
Lab in the original box to transport
the NitroFlow
®
Lab over longer distances.
Lift the NitroFlow
®
Lab with a forklift.
For qualifications of personnel, refer to §2.1.
5.2 Define location
IMPORTANT
The NitroFlow
®
Lab contains compressors that
generate heat; for optimal performance and lifetime it
is necessary that cooling air can be vented without
resistance. A minimum clearance distance from walls
or other objects of at least 50 cm/ 20 inches on all
sides (back, left, right and top) is a necessity; also
efficient local ventilation at the ventilation outlet is
highly recommended especially when the device is
installed under a bench
Install the NitroFlow
®
Lab on a fixed location. The location must meet the
following requirements:
Minimum clearance of 50 cm on all sides (back, left, right and
top) as to facilitate heat removal
Indoors
Dry
No continuous direct irradiation by sunlight
Away from heat sources
Properly ventilated room.
Easy accessibly for operating and service