K3.1.142e Manual NitroFlow Lab - 41 -
7.5 Stop NitroFlow
®
Lab
1. Press the ON/OFF button to switch the unit OFF (in case it is
operating).
2. Switch off the power switch before you perform maintenance.
3. Make sure the system is depressurized; check the internal pressure
level in the maintenance menu
(screen 3/5)
4. When you restart afterwards there is a 3 minutes delay before it starts
again.