K3.1.142e Manual NitroFlow Lab - 33 -
6.3.6 Maintenance menu
Access: Touch maintenance menu button in settings screen
(refer to § 6.3)
Function: Shows maintenance status and offers calibration
possibility.
The maintenance menu consists of 5 different screens. Each screen
displays maintenance status or calibration buttons.
SCREEN 1/5
Data Explanation
Type Shows type of generator this unit is
Version Software revision number
O2 lifetime Month-year when O
2
-sensor needs to be changed (3
years from data of order)
Filter life Hours countdown from 1 year to 0 hrs