K3.1.142e Manual NitroFlow Lab - 28 -
6.3.2 Alarm settings menu
Access: Touch alarm settings menu button in settings screen
(see § 6.3.2)
Function: Set different alarms
In the alarm settings menu it is possible to set 6 different alarms.
Screen Alarm Explanation
1/6 O2 high oxygen level too high
2/6 O2 low oxygen level too low
3/6 Pres. Inlet high inlet pressure too high
4/6 Pres. Inlet high inlet pressure too high
5/6 Pres. Outlet high outlet pressure too high
6/6 Pres. Outlet low outlet pressure too low
1. To activate an alarm touch button A. When the button is touched
you can select the options YES, AUTO RESET or NO by pressing
the arrow keys
Default all alarms are set to NO, which means they are not
activated; activating the alarms or not is the choice of the
user; alarms do not influence the output and purity.
2. When you select YES or AUTO RESET, the rest of the alarm
parameters that need to be set will pop-up automatically (see
screen below).
A