K3.1.142e Manual NitroFlow Lab - 42 -
8 Troubleshooting
8.1 Error list
Error Possible cause Possible solution
Main switch is off Switch main ON and push
power switch ON.
No start and
no display
No power to supply outlet Check electrical panel circuit
breaker.
Ambient temperature is too
high
Lower the temperature, if
possible
Check whether the minimum
clearance between the
NitroFlow
®
Lab and the walls
is large enough.
Inlet carbon adsorber filter is
polluted
Replace the inlet filter
NitroFlow
®
Lab is switched
off
Switch on the NitroFlow
®
Lab
Leak in piping Check for leaks in the piping.
Nitrogen outlet line is
blocked.
Check/open the outlet line
Delivery of
nitrogen too
low or absent
Temperature is too high
A.3.
Ambient temperature is too
high (over 35°C/ 95°F).
In-/outlet gratings are
clogged.
Cooling fans are not or
insufficiently functioning.
Compressors are
overloaded.
Pressure in nitrogen storage
vessel over 8 bar(g) because
of erroneous setting of
pressure switch
Reset pressure switch levels
Residual
oxygen
content too
high
Ambient temperature lower
than normal
Increase temperature or
readjust purity (refer to §7.3)