K3.1.142e Manual NitroFlow Lab - 21 -
5.3 Unpack and check equipment
Open the packaging per instructions on the crate.
Make sure that all components are delivered. Refer to § 4.1.
5.4 Connect nitrogen consumer
Warning
Do not connect the power at this time.
Make sure that the inlet and outlet tubes are free of
dust, particles, metal parts and curls, liquids and
grease before you connect the NitroFlow
®
Lab.
Connect the product outlet to the application.
5.5 Connecting power
1. Connect the mains plug to a suitable wall socket with earth connection
2. The control system has in- and output contacts for remote control and
alarm signaling (refer to §5.6).
Warning
The main supply line voltage must be within 10% of
nominal rated voltage for the generator. In case of larger
variations the NitroFlow Lab will stop; continued use
under these circumstances will inevitably lead to motor
damage.