K3.1.142e Manual NitroFlow Lab - 29 -
Button Selection Result
Active No Alarm function for this parameter is not active
Active Yes Alarm function for this parameter is active;
alarm messages must be reset manually
Active Auto reset Alarm function for this parameter is active;
When alarm level is not exceeded any longer
before manual reset, the alarm will reset itself
Stop Yes Generator will switch off in case alarm level
is exceeded
Stop No An alarm signal will be given but generator
will continue to run in case alarm level is
exceeded
Level 0-16% O
2
100 – 84% N
2
For screen 1/6 and 2/6: this is the oxygen- or
nitrogen level* at which the alarm is set.
Level 0-13 BAR*
0-188.5 PSI*
For screen 3/6 and 4/6.
This is the pressure level at which the alarm
is set
Level 0-10 BAR*
0-145 PSI*
For screen 5/6 and 6/6.
This is the pressure level at which the alarm
will appear.
Delay 0-300 sec Delay time in seconds between the moment
that the alarm level has been exceeded and
signaling; this feature prevents false alarms
in case of short spikes
*see also local settings -menu
ATTENTION:
It is impossible to set O
2
low at a higher level than O
2
high. The setting of O
2
-low is limited once O
2
-high has
been set. Therefore first set O
2
high level before setting
the O
2
low level.