K3.1.142e Manual NitroFlow Lab - 49 -
6. Use a 13mm spanner to unscrew the four bolts holding the
compressors to the unit on both sides.(Fig. 9-12)
7. Use a 10mm spanner to push the ring in with the plastic tube. Hold the
spanner while pulling away the tube. (Fig. 9-13)
8. Use an 8mm spanner to push ring in with the pipe. Hold the spanner
while pulling away the pipe (Fig. 9-14)
9. Disconnect the multi-pin plug from the socket by squeezing the clips on
each side. (Fig. 9-15)
10. Lift and withdraw the compressor assembly from the unit (Fig. 9-16).
Fig. 9-12 – unscrew bolts
from compressors
Fig. 9-13 – pull away pipe
Fig. 9-14 – pull away pipe
Fig. 9-10 – remove
second cover
Fig. 9-11 undo bull nose
connections
Fig. 9-9 – unscrew side
panels