The TRUE-IPV® In-Line Valve is intended to be used to provide IPV® (Intrapulmonary
Percussive Ventilation) therapy to intubated patients while assisted by Conventional
Mandatory Ventilation (CMV) using pressure-control, volume-control, SIMV-PC etc.,
when direct connection of IPV® is not indicated.
TRUE-IPV® in-line use with a ventilator is based on patient need, from 2 times per day up
to 6 times per day (every 4 hours), as recommended by physician. Always use institutional/
hospital protocol when possible.
WARNING: Follow institutional protocols before disconnecting ventilator
inspiratory limb prior to installation of Percussionaire® TRUE-IPV® In-Line Valve.
Chapter 3: TRUE-IPV® In-Line Valve with a Ventilator
TRUE-IPV® In-Line Valve
22mm I.D. x 22mm O.D.
Single Patient
Intended Use
Treatment Frequency
TRUE-IPV® In-Line Valve
NOTE: For use only with Percussionaire® TRUE-IPV® ventilator devices. Compatible with all
Percussionaire® single patient Phasitrons. The IPV® In-Line Valve is indicated for neonatal,
pediatric, and adult patient populations, for whom IPV® therapy has been prescribed.
Pressure relief adjustment valve
Phasitron® 5 port
Pressure relief port
22mm ID
22mm OD
11