34
Administering TRUE-IPV® Therapy with In-Line Valve
When administering TRUE-IPV® Therapy with an In-Line Valve, a pressure control (PC)
mode is recommended or follow your institutional protocol.
Mean Airway Pressures will increase slightly with the administration of TRUE-IPV®
in-line therapy with the ventilator. The respiratory care practitioner must be aware
of this eect and monitor the patient closely for any adverse side eects.
When using the IPV®-2C in pressure-control mode, the In-Line Valve may remain
closed. When using the ventilator in volume-control, the In-Line Valve may be opened
to create a leak.
WARNING: NEVER run device without liquid in nebulizer during treatment.
This is required for airway hydration.
WARNING: Note the current ventilator alarm and mode settings.
WARNING: Reset occasional CMV high-pressure alarms as they occur. When
applying TRUE-IPV® in-line, adjust the pressure relief valve to achieve desired
Amplitude Pressure per your institutional/hospital protocol. High-pressure alarms
should not occur on a regular basis if the Pressure Relief Adjustment valve is set correctly.
NOTE: Patients who are performing T-tube trials or CPAP sprinting may be taken o the
ventilator for the IPV® treatment utilizing a ex adapter. Decreasing cu pressure still
applies to this patient population.
NOTE: Lowering of the cu pressure facilitates secretion removal into the oral cavity
where they may be suctioned. This also helps in the prevention of tube obstruction in
the event copious secretions are mobilized.
NOTE: Following your institutional protocols for cued endotracheal tubed patient,
the cu pressure may be lowered.
1. Ensure the IPV®-2C is “OFF” and connected to a 50 psi/3.2 bar gas source.
2. Connect the IPV®-2C to 50-80 psig (345-551 kPa) gas power source.
Master Switch is “OFF”.
3. Put prescribed medications into nebulizer and add diluent as directed by
physician to a maximum of 20 ml. If no medications are prescribed, use
normal saline or sterile water, as directed by physician. Nebulizer should
always be on.
4. With IPV®-2C Master switch “ON”, rotate Operational Pressure control knob
for an operating pressure of 35 to 40 psig (206-275 kPa).
5. Rotate Inspiratory Time and Frequency knobs to the 9:00 position
which produces an approximate Pulse Frequency rate between 300-350.
The i:e ratio will be approximately 1:1 (may be used for recruiting alveoli).
6. Turn on Inspiratory Flow control knob slowly adjust for desired
chest wiggle and amplitude.
(Continued on p.35)