38
Cleaning Phasitron®5
1. Thoroughly rinse each of the disassembled parts (except for tubing harness
and lter) under warm running tap water for approximately 10 seconds.
2. Use fragrance-free liquid soap added to a clean bowl or basin with
warm water.
3. Hand wash all parts of the Phasitron®5 kit and accessories in the warm
soapy water.
4. Rinse all parts thoroughly using sterile water.
5. Gently shake all parts to remove as much water as possible, then dry with
a clean lint-free cloth or paper towel.
6. Use a clean, damp cloth to wipe the exterior of the tubing harness with an
approved alcohol based cleaner.
7. Reassemble the Phasitron®5 and place in the supplied bag until next use.
8. Do not disinfect the Phasitron®5 for reuse with more than one patient.
Chemical Class Active Ingredient
Bleach 5.25% Sodium hypochlorite
Alcohol 70% Isopropyl alcohol
Peroxide 3% Hydrogen peroxide
Benzyl Ammonium N-alkyl dimethyl ethyl benzyl ammonium chlorides
Chloride N-alkyl dimethyl benzyl ammonium chloride
Phenolic Ortho-phenylphenol
Ortho-benzyl-para-chlorophenol
Quaternary Ammonium Didecyl dimethyl ammonium chloride
Chloride Alkyl dimethyl benzyl ammonium chloride
Cleaning and Disinfecting Solutions
NOTE: The Phasitron®5 breathing circuit has been tested for biocompatibility with
the following cleaning and disinfecting solutions: