Disassembly of the Phasitron®5
Disconnect tubing from IPV®-2C device and Phasitron®5.
1. Gently rotate nebulizer bowl towards
the rear of the Phasitron®5 until it stops.
2. Gently separate nebulizer from
Phasitron®5.
3. Holding the nebulizer cap, twist the
nebulizer bowl to remove bowl from cap.
Discard any unused medication in
accordance with hospital/institutional
protocol.
4. Twist white cap on the rear of
Phasitron®5 to remove.
5. Remove cap.
6. Remove sliding venturi with spring
from Phasitron®5 body.