14
The patented Phasitron®5 uses a unique venturi as a “clutch” mechanism to protect the
lung from over pressure. By automatically adjusting to the resistance of the lung, the
Phasitron®5 precisely and safely delivers the optimal amount and ow of air required by the
alveolar space. When lung resistance is low, as in a compliant lung, all the pulsed air from
the IPV®-2C device enters the mouth of the venturi. Each air pulse draws up to four times as
much additional air into the venturi tube. This low-pressure entrained air automatically lls
the available space in the lung. The Phasitron®5 continuously and instantaneously adjusts
to keep a gentle and safe air pressure, even in a compromised lung.
NOTE: TRUE-IPV® therapy can only be achieved using the Phasitron® 5.
Phasitron®5 Breathing Circuit Setup