20
Knob Functions
DEMAND CPAP/PEEP
INSP TIME
INSP FLOW
FREQUENCY
MASTER
Chapter 5: Controller Functions
Knob, Switch and Button
Demand Continuous Positive Airway Pressure
(Demand-CPAP) for reducing the work of breathing
Lifts the baseline in a static way
Aects peak and mean airway pressure
Used to reduce work of breathing (WOB),
stabilize upper airways
The Inspiratory Time control knob arrow when
rotated (counterclockwise) to
the 12:00 default position increases the
Inspiratory Time while simultaneously
decreasing the Expiratory Time.
Sub-tidal inspiratory/expiratory ratio
Ranges from 1:1 to 1:3
Aects mean airway pressure
The Inspiratory Flow control knob determines
the amplitude delivered to the patient during
inspiratory time.
Set according to patients need or tolerance.
Chest wiggle
The Frequency knob controls the rate of high
frequency volumes delivered.
Ranges from 60 to 330 cycles per
minute
Aects peak and mean airway pressure
The Master Switch turns the IPV®-2C controller
ON and OFF.
The patient receives TRUE-IPV® treatment when
the Master Switch is in the ON position.