Knob Function
The Manual Inspiration button controls the
delivery of a regulated source of gas through the
orice of the Phasitron®5 venturi.
To remove CO
2
To induce a cough
WARNING: The longer the Manual
Inspiration button is pressed, the greater the
potential for tidal volume delivery.
WARNING: NOT FOR USE WITH NEONATES.
The Nebulizer ON/OFF Switch turns the aerosol
delivery ON and OFF.
Recommended to always use nebulization
The Operational Pressure knob controls the
peak operating pressure of the entire unit.
The optimal pressure is 40 psig (3.4 bar, 345 kPa)
for Pediatrics and Adults and 30 psig (2.07 bar,
207 kPa) for Neonates.