PerkinElmer OPERA PHENIX PLUS - Confirmation Form and Network Options

PerkinElmer OPERA PHENIX PLUS
46 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Send Back (Fax or Email)
http://www.perkinelmer.com/AboutUs/ContactUs/ContactUs
Date:
To: PerkinElmer
Fax Number:
Customer:
Subject: Installation of Opera Phenix Plus
I have reviewed the Site Readiness Instructions and confirm that:
Meets Requirements
o
The available space at the installation location meets the space requirements
of the Opera Phenix Plus (including upgrade components, if applicable). See
chapter 2.3 .
o
Additional space for unpacking/installing the Opera Phenix Plus is available as
specified at the installation location (only during delivery). See chapter 2.3 .
o
The laboratory table has a carrying capacity of at least 400 kg (882 lb).
o
The laboratory meets the temperature (20 °C to 25 °C) and humidity (less than
70 % non condensing) requirements.
o
The room is well ventilated and free from dust.
o
At least three electrical outlets are available within 1.5 m (59 in) of the
installation location.
o
Electrical supply has a frequency of 50 or 60 Hz and a voltage in the range of
100 to 120 V or 220 to 240 V.
o*
* Only with optional ECO upgrade: CO
2
supply is provided by the customer as
specified in the gas requirements. See chapter 2.7 .
o
The internal transportation of the shipment containers from the customer's
receiving department right to its desired room position is guaranteed.
o
Any customer-supplied PC for Harmony office installation meets the
requirements described. See chapter 3.3 .
o
The packing material will be disposed of by the customer.
o
The local IT department has agreed to the integration of the Opera Phenix Plus
system (including the Harmony PC running with Windows
®
10).
The preferred solution for the network integration will be:
o No integration
o Simple Integration (chapter 3.1.2 ) o Office PC big (chapter 3.1.3.2 )
o Office PC small (chapter 3.1.3.1 ) o Router (chapter 3.1.3.3 )
o
The required hardware for the network integration will meet the requirements
and will be available for the installation. See chapter 3.1 .
Sincerely,
Name
Telephone Number
4 Confirmation Fax/Email
44 Opera Phenix Plus Site Readiness