Electromagnetic Compatibility
B-6 Efficia CM Series Service Guide
Table B-6 Patient Coupling Port
Immunity Test Immunity Test Level Compliance Level
Electromagnetic
Environment - Guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
±8 kV contact
±2, ±4, ±8, ±15 kV air
±8 kV contact
±2, ±4, ±8, ±15 kV air
Floors should be wood,
concrete, or ceramic tile. If
floors are covered with
synthetic material, the
relative humidity should
be at least 30%.
Conducted disturbances
induced by RF fields
IEC 61000-4-6
3 V, 0.15 MHz – 80 MHz
6 V in ISM and amateur
radio bands between
0.15 MHz and 80 MHz
3 V, 0.15 MHz – 80 MHz
6 V in ISM and amateur
radio bands between
0.15 MHz and 80 MHz
Typical Hospital
Environment