Test Recording
Performance Verification Testing
3-4 Efficia CM Series Service Guide
Test Recording
Authorized Philips personnel report test results to Philips to add to the product
development database. Hospital personnel, however, do not need to report results.
The following table describes what to record on the service record after you complete
the tests in this chapter.
Note — P = pass, F = fail, X = measured value as defined in tests in this chapter.
“Cardiac Output Test (C.O.)” on
page 3-24
• Patient simulator
• Fluke PS420 or similar M1012-61601 test
tool
• Fluke cable CD 2462364 and M1642A or
similar
“Safety Tests” on page 3-25 A multimeter with a leakage network or a
product safety analyzer
“Nurse Call Relay Test” on
page 3-29
• A patient simulator
• An ohmmeter
• A phono connector
“Network Test” on page 3-31 None
To Perform This Test You Need This Test Equipment
Test What to record
Visual V:P or V:F
Power-On PO:P or PO:F
Alarms AL:P or AL:F
SpO
2
SpO
2
:P or SpO
2
:F
ECG ECG:P or ECG:F
Respiration R:P or R:F
ECG Output EO:P or EO:F
NBP NBP:P/X1/X2/X3 or
NBP:F/X1/X2/X3
IBP IBP(1):P or IBP(1):F
IBP(2):P or IBP(2):F
Sidestream CO
2
SCO
2
:P or SCO
2
:F
Mainstream CO
2
MCO
2
cal:P or MCO
2
cal:F
C.O. C.O.:P orC.O.:F