EasyManua.ls Logo

Philips Efficia CM Series - Temperature Test; Cardiac Output Test (C.O.)

Philips Efficia CM Series
116 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Temperature Test
Performance Verification Testing
3-24 Efficia CM Series Service Guide
To test the ECG output:
Temperature Test
To perform this test, you need a patient simulator (with 0.1
o
C or 0.2
o
F). Perform the
following test for each temperature connector in use.
To test the performance of the temperature measurement:
Cardiac Output Test (C.O.)
This test checks the performance of the C.O. module.
To perform this test, Philips recommends that you use the following equipment or its
equivalent:
Fluke PS420, M1012-61601, CD 2462364 or similar
Step
1
Connect the patient simulator to the ECG connector on the monitor and
the defibrillator to the ECG output on the monitor.
2
Set the patient simulator to the following configuration:
HR = 100 bpm
ECG sinus rhythm
3
Turn on the defibrillator in simulation mode.
4
Verify that the ECG signal is displayed.
5
If you do not get the expected results, see “ECG/Respiration Measurement
Problems” on page 4-9.
Note — Philips employees record this value as EO:P or EO:F.
Step
1
Connect the patient simulator to the temperature connector on the
monitor or measurement server extension.
2
Configure the patient simulator to 40
o
C (100
o
F).
3
Wait for the monitor to display the static temperature value.
The value should be 40
o
C ± 0.2
o
C (100
o
F ± 0.4
o
F).
4
If the value is not within tolerance, see “Temperature Measurement
Problems” on page 4-7.
Note — Philips employees record this value as T(1 or 2):P or T(1 or 2:F)

Table of Contents

Other manuals for Philips Efficia CM Series

Related product manuals