Troubleshooting Tables
Troubleshooting
4-4 Efficia CM Series Service Guide
Troubleshooting Tables
Use the following tables to diagnose and fix monitor problems. The tables describe a
monitor problem by symptom, list a possible cause, and suggest actions. Check the
first possible cause listed, and then perform the associated action to repair the
problem. Perform all actions in the order that they appear in a table. If you cannot
resolve the problem by following the actions in the table, contact your service provider
for repair.
Monitor problems are categorized as follows:
Power
Display
Alarms
Measurements, including:
NBP
Temperature
SpO
2
–CO
2
ECG/Respiration
IBP
C.O.
Navigation wheel, touch screen, and keys
Recorder
Nurse Call
Security
SD Card
If using the following troubleshooting tables does not solve the problem, see “Running
System Diagnostics” on page 4-13.
Press the On/Standby key
when the monitor is in Standby
mode (with AC power and/or
battery power).
A tone sounds and the Philips splash screen
appears.
If configured, the Admit Patient Menu appears.
After you acknowledge the Admit Patient Menu,
the main screen appears.
Press the On/Standby key
when the monitor is on (with AC
power and/or battery power).
The Philips splash screen appears and the LCD
turns off.
Select
Shut Down on the
General tab in the System Menu.
After you select OK, the Philips splash screen
appears while data is being saved. The LCD turns
off.
User Action Expected Result