Performance Verification Testing
Efficia CM Series Service Guide 3-5
Accessing the Admin Functions
Accessing the Admin Functions
You use the System Menu to access the password-protected Admin functions, which
allow you to configure the monitor and perform diagnostic and maintenance tasks. For
more information about configuring the monitor, see the Efficia CM Series Patient
Monitors Instructions for Use or the Efficia CM Series Patient Monitors Installation and
Configuration Guide.
Notes
The Admin tab in the System Menu is password-protected, and should only
be accessed by qualified service personnel.
The procedures in this guide use the default Administrator password, 215. To
improve the security of the monitor settings, you can change the
administrator password after you install the monitor. For more information
about changing the password, see the Efficia CM Series Patient Monitors
Installation and Configuration Guide.
Caution The System Menu fills the entire monitor display when it is open and obscures any
active measurement data. When the Admin tab is open, it does not automatically
close if left unattended, but remains open until it is closed manually.
To access the System Menu and the Admin tab:
Temperature T(1):P or T(1):F
T(2):P or T(2):F
Safety S(1):P/X1/X2 or S(1):F/X1/X2
S(2): P/X1 or S(2): F/X1
S(3): P/X1 or S(3): F/X1
Nurse Call NC:P or NC:F
Network Test NT:P or NT:F
Test What to record
Step
1
Select System ().
The System Menu appears.
1
Select the Admin tab.
2
In the window that appears, select the password field.
A keyboard opens.