Running the Self Test
Troubleshooting
4-14 Efficia CM Series Service Guide
Running the Self Test
Running the Self Test performs the resident self test, which includes a complete
memory test and a front end test on all available modules, such as SpO
2
or NBP. If the
test cannot detect a specific module or if an error occurs, then that test fails.
Note — When the monitor is powered up from Standby mode, the monitor performs the
front end tests in addition to the start-up sequence. When the monitor is
powered up from Deep Sleep mode, the monitor performs the complete
memory test and front end tests.
To run the self test:
Testing the Recorder
To test the optional recorder:
Step
1
In the Diagnostics Menu, select the Self Test button.
The following messages are displayed as the Result for each Test Item:
Passed — the module is installed, passed the self test, and is operating
correctly.
In Progress — the test is running.
Not Supported — the module is not installed or configured.
Failed (in red text) — the test was not successful.
2
When all tests are complete, close the menu.
Detected errors are written to the error log with detailed error information
and error codes.
Step
1
In the Diagnostics Menu, select the Recorder Test button.