Troubleshooting Tables
Troubleshooting
4-10 Efficia CM Series Service Guide
IBP Problems
Symptom Possible Cause Action
IBP measurements are not
displayed.
The transducer malfunctioned. Contact your service provider.
IBP measurements are
unreliable.
The transducer is not zeroed. Refer to your facility’s protocol
regarding pressure monitoring
setup and transducer
placement.
Perform a zero. See the Efficia
CM Series Patient Monitors
Instructions for Use.
The transducer cannot be
zeroed.
Ensure the correct steps for
performing a zero (especially
the position of the stopcocks)
are followed. See the Efficia
CM Series Patient Monitors
Instructions for Use.
Replace the transducer cable.
Replace the transducer tubing
setup.