Visual Test
Performance Verification Testing
3-8 Efficia CM Series Service Guide
Demo mode is used to demonstrate the monitor without connecting the monitor to an
actual person. Demo mode simulates all patient parameters and generates alarms
when alarm settings are exceeded.
Caution Entering Demo mode clears the patient data.
To put the monitor in Demo mode:
Visual Test
To perform the visual test:
Step
1
Access the Admin tab. See “Accessing the Admin Functions” on page 3-5.
2
Select Demo Mode.
Note — Checking Demo Rack allows you to demonstrate the rack
measurements when Demo Mode is enabled.
3
Close the menu.
A confirmation window appears.
4
Select OK.
The monitor enters Demo mode and clears all patient data. DEMO
appears on the top of the monitor display.
5
To exit Demo mode, press the On/Standby key to turn the monitor off.
The monitor clears all simulated patient data.
Step
1
Inspect the system for obvious signs of damage; for example, cracks, cuts,
or breakage.
2
Check all external cables and accessories for damage; for example, cuts,
kinks, wrong connections.
3
Ensure that all markings and labeling are legible.
If the labels on the rear case are not legible, replace the rear case. If the
serial number label is not legible, contact the Philips Customer Care
Center or your local Philips representative to return the monitor for label
replacement.
4
Check for any obstructions to mechanical parts.
The expected test result is that the system has no obvious signs of
damage or obstruction.
Note — Philips employees record this value as V:P or V:F.